顶部链接&语言切换

Time of issue:2021-09-16 17:43:45

Home  |  About us  |  Products  |  News  | Resoures  |  Careers  |  Contact us        CNEN

imgboxbg

Products

SARS-CoV-2 Antigen Rapid Test Kit(Colloidal gold)
1/1
Number of views:
1000
The PocRoc® SARS-CoV-2 Antigen Rapid Test Kit(Colloidal gold)is used for qualitative detection of SARS-CoV-2 virus nucleocapsid protein, which is an important conserved structural protein of SARS-CoV-2 in human nasal/ oropharyngeal samples. The detection of SARS-CoV-2 nucleocapsid protein antigen can be used to assist the diagnosis of novel coronavirus infection, and is helpful for the early detection of novel coronavirus pneumonia infection in the latent period. For Medical Professional Use Only!
Number of views:
1000
Product serial number
Quantity
-
+
1
General Info
Video

Intended use

The PocRoc® SARS-CoV-2 Antigen Rapid Test Kit(Colloidal gold)is used for qualitative detection of SARS-CoV-2 virus nucleocapsid protein, which is an important conserved structural protein of SARS-CoV-2 in human nasal/ oropharyngeal samples. The detection of SARS-CoV-2 nucleocapsid protein antigen can be used to assist the diagnosis of novel coronavirus infection, and is helpful for the early detection of novel coronavirus pneumonia infection in the latent period.

For Medical Professional Use Only!

 

Product Information

Production name

Test samples

Specifications

PocRoc ® SARS-CoV-2 Antigen Rapid Test Kit(Colloidal gold)

human nasal/ oropharyngeal samples

 

1T/Kit, 5T/Kit, 25T/Kit

 

Storage Conditions

All reagents are ready to use as supplied. Unopened reagent kits are stable at 4°C ~30°C for 24 months tentatively. It should be used within 1 hour once the pouch is opened. Do not freeze the kit or expose the kit above 37°C during storage.

Test procedure

1、Open the package and take out the test card.

2、Place the extraction tube on the workbench. The swab extractor bottle is pressed vertically downward to the extraction tube.

3、Put the swab specimen into the extraction tube, rotate the swab for about 10 seconds, and press the swab head against the tube wall to release the antigen in the swab. Squeeze the swab over the head to remove the swab so as to remove as much liquid as possible from the swab. Dispose of swabs according to biohazard waste disposal method.

4、Install the beater on the extraction tube, put two drops into the sample hole of the test card, and start the timer.

Notes: Applying sufficient amount of sample extraction liquid is essential for a valid test result. If migration (the wetting of membrane) is not observed in the test window after one minute, add one more drop in the extraction tube to sample well.

5、Read the results in 20 minutes. A strong positive result can be reported within 20 minutes, but a negative result must be reported after 20 minutes, and the result after 30 minutes is no longer valid.

We could not find any corresponding parameters, please add them to the properties table

底部版权信息

Time of issue:2021-06-08 11:15:00

Copyright©2021 Lumigenex (Suzhou) Co., Ltd. All Rights Reserved. Website construction: www.300.cn

底部备案信息

Time of issue:2021-09-17 09:20:58