Products
Myoglobin Test is used for rapid diagnosis at early stage of acute myocardial infarction.
Intended Use
This Myoglobin (MYO) Test Kit (Time Resolved Fluorescence Immunochromatographic Assay) is suitable for quantitative detection of Myoglobin(abbreviated as MYO below)in human Serum/plasma/whole blood in vitro.
Product Information
Production name |
Test samples |
Specifications |
MYO Quantitative Test Kit (TRFIA) |
serum/plasma/whole blood |
20T/Kit, 50T/Kit |
Storage Conditions and Expiry Date
The validity of test kit sealed in an aluminum foil pouch is 18 months under condition of 2-8℃. The sealed test kit is stable for up to 30 days under room temperature (2-30℃) within validity period. Temperature change (< 37℃) during shipment has no effect on quality of the test kit. The unsealed kit is stable within 1 hour under ambient condition.
Product performance
Accuracy: the relative deviation from the target value of the enterprise reference should be within ±15% .
Limit of detection: no more than 1.0 ng/mL.
Linear range :5~400 ng/mL range, the kit correlation coefficient r≥0.9900.
Repeatability: coefficient of variation (CV)≤15%.
Inter-batch difference: coefficient of variation (CV)≤15% between batches
Stability: test the physical properties, accuracy, minimum detection limit, linear range and repeatability of the kit within one month of service life,Shall comply with the corresponding requirements.